• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVHD-DF16
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2024
Event Type  malfunction  
Event Description
During an atrial tachycardia procedure, there was a communication issue with the advisor hd mapping catheter resulting in a delay.An error message appeared on the ensite x system that read, "catheter in port 4 is invalid because it has already been used with a different patient.Disconnect it and try another catheter.¿ despite using a new catheter.The connecter was changed, the amplifier and ensite x dws were restarted, but the issue persisted.The advisor hd mapping catheter was exchanged, and the procedure was completed with no adverse consequences to the patient.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported communication issue and subsequent delay could not be conclusively determined.
 
Manufacturer Narrative
Additional information was received confirming the incorrect lot number was originally provided.The correct lot number of the complaint device is 8981801.Additionally, the manufacturer of lot number 8981801 is plymouth mfr 3005334138.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key19140997
MDR Text Key340567760
Report Number3008452825-2024-00230
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public(01)05415067028198(10)10208592(17)270131
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K202066
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD-AVHD-DF16
Device Lot Number10208592
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/18/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/17/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/15/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-