• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX Back to Search Results
Model Number 866389
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2023
Event Type  malfunction  
Manufacturer Narrative
It is unclear with the information available if this is a reportable event.Based on the information received, the reported issues were not life-threatening, there was no permanent damage or permanent impairment to the patient, and medical intervention was not required; therefore, this event does not meet criteria for serious injury.Some of the information provided remains unclear as to the system¿s behavior; however, there do appear to be conflicts encountered by the user during admit, discharge and transfer (adt) activities.Adt data conflicts are presented to the user when they occur with options for resolving the data conflict.The user is prompted to make a selection and an icon appears to alert the user to the issue.Product labeling describes data conflict behavior and alarming continues during a conflict.A yellow alarm at the intellivue patient monitor is issued periodically if the conflict message is not addressed by the clinician.In addition, the user explained they were aware of the adt issues at the time of the event, so the adt issues were not likely to result in patient harm.A philips clinical application specialist has reached out to the hospital and had a dialogue with the person responsible for the implemented piic ix solution to provide workflow information/explanation and application process support in current setup to minimize the risk for similar events to happen in the future.Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.E1; reporting institution phone +(b)(6).E1; reporter phone +(b)(6).
 
Event Description
User report (b)(4) was received which reported the customer experienced adt data conflicts and they indicated has happened at least 3 times almost identically.
 
Manufacturer Narrative
Based on the information available in the case and provided by the philips clinical application specialist (cas), the root cause of the issue was confirmed to be an operator workflow issue.The philips cas has reached out to hospital and had a dialogue with the responsible department for the implementation of the piic ix solution.In addition, the hospital was provided workflow information/explanation and application process to support the current setup, to minimize the risk for similar events that may happen in the future.Supplementary training on workflows have been proposed to the customer for the impacted departments in hospital, though not yet implemented.No further investigation or action is warranted at this time.H3 other text : provided information for workflow solution and training.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT INFORMATION CENTER IX
Type of Device
PATIENT INFORMATION CENTER IX
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
deborah currlin
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key19141272
MDR Text Key341473833
Report Number1218950-2024-00276
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K211900
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Device Catalogue Number866389
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/19/2024
Supplement Dates Manufacturer Received04/10/2024
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-