Device evaluated by manufacturer: the device was returned for analysis.The first photo shows that the tip had slight bends and was stretched with a partial detachment at the end of the tip by the sensor housing.The second photo is the fluro and showed the wire, but you couldn't tell from the photo if the wire was damaged.The third photo showed no pd value on the polaris ilab screen.The shroud of the occ cable was connected to the bench top testing equipment.The rfid tag associated with this device is e004010841e7ea38 and the coefficient values were confirmed to be programmed.Visual inspection revealed that here was no tip attached to the wire.The shroud of the occ cable was connected to the ffr link for signal verification.The signal was not present, as designed.During testing with the ffr link, the 'signal strength' led showed a yellow/orange light and the 'zeroed' led showed a blinking red light, indicating that the wire was not working appropriately.The led lights for a properly working wire are green.The shroud of the occ cable was connected to the bench top testing equipment.The modulation was checked but there was no modulation (0%) for this device.Microscope inspection revealed that the tip was detached at the sensor housing.Inspected the sensor by viewing the sensor port to verify that the sensor was in the correct location.The sensor appeared to be inside the sensor housing detached from the fiber optic cable.The wire was gently shaken to see if the sensor would move within the sensor housing and the sensor did move which verifies the sensor was detached from the fiber optic.The proximal end face of the wire showed the proximal end face was cracked/damaged.When the handle was dismantled, it was found that the proximal end of the wire appears to have punctured the polymer disc.With the proximal end of the wire damaged, the handle was dismantled to see if there was any damage or irregularities.
|