Corrected fields: h6 (medical device problem code).This report is being supplemented to correct information provided in the initial medwatch report and to provide additional information, based on the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, it is likely, that an image was not displayed, due to a contact failure of the lvds (low voltage switching device unit).However, a conclusive root cause could not be determined.Olympus will continue to monitor field performance for this device.
|