Model Number LF-TP
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed during the device inspection, that the tracheal intubation fiberscope coating of the insertion tube was peeled off.There were no reports of patient involvement.
 
Manufacturer Narrative
The device evaluation found that the connecting tube was deformed.The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of observed insertion tube coating peeling issue could not be determined, however, the issue was likely due to physical impact stress and or chemical stress due not reprocessing in accordance with the ifu.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACHEAL INTUBATION FIBERSCOPE
Type of Device
TRACHEAL INTUBATION FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19144836
MDR Text Key341469307
Report Number9610595-2024-08290
Device Sequence Number1
Product Code EOQ
UDI-Device Identifier04953170136825
UDI-Public04953170136825
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K981543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLF-TP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/19/2024
Supplement Dates Manufacturer Received05/20/2024
Supplement Dates FDA Received05/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/12/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-