It was reported that the catheter shaft was frozen.During a pulmonary vein isolation (pvi) ablation procedure, a polarxfit catheter was selected for use.After having prepared polarxfit balloon properly, soon after ablation started, temperature did not decrease, and physician noticed that the catheter handle was getting cooler.Ablation was stopped after a few seconds and noticed that gas cable connection was frozen, probably because some water was present during preparation and it was not noticed.Then, it was noticed that the polarmap was trapped in the balloon, when pulling it outside the patient, it was noticed that the catheter was frozen.The handle and the proximal portion of the shaft were frozen.Finally, it was decided to replace the polarxfit and the polarmap and the issue was resolved.Procedure was completed successfully without any patient complications.Catheter is being returned for further analysis.
|