• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number 2C9961KP
Device Problem Filling Problem (1233)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2024
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a multirate large volume infusor underinfused during infusion.After the expected therapy time was complete, it was observed that the infusion had not finished and therefore was lfor for three more hours to finish the medication delivery.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The lot was manufactured between august 31, 2023 and september 1, 2023.The actual device was received for evaluation.A visual inspection was performed which did not identify any abnormalities that could have contributed to the reported condition.A functional flow rate test was performed, and results were within the product specification range.The reported condition was not verified.The device was determined to be conforming product.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19146964
MDR Text Key341367867
Report Number1416980-2024-01816
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412081502
UDI-Public(01)00085412081502
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K011317
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C9961KP
Device Lot Number23H021
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/19/2024
Supplement Dates Manufacturer Received05/21/2024
Supplement Dates FDA Received05/24/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNSPECIFIED MEDICATION.
-
-