SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE Back to Search Results
Model Number UM-G20-29R
Device Problems Break (1069); No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/19/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found indentation at the tip of the insertion and damage to the ultrasonic probe.A device history review revealed no issues that could have caused or contributed to the reported issue.This issue is addressed in the instructions for use (ifu): 3.5 how to use this product (warning) ¿ do not insert the ultrasound probe into the endoscope without a clear view of the endoscope.Doing so may result in damage to tissues in the body cavity and damage to the product.¦observation of ultrasound images (warning) ¿ do not push or pull the ultrasonic probe with strong force or pull it into the endoscope's curved section while the probe is driven (unfrozen state).Push and pull the ultrasonic probe slowly, especially when the endoscope is strongly bent or forceps is being raised.Strong force or sudden movements may cause image distortion or damage to the ultrasound probe.¿ when driving the ultrasound probe, return the endoscope curvature and forceps raising as much as possible.Driving with a harsh probe geometry may lead to image distortion or damage to the ultrasound probe." olympus will continue to monitor the field performance of this device.
 
Event Description
It was reported that the subject device had image display issue.The issue occurred during preparation for use.The procedure was completed using similar device.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19149116
MDR Text Key340952369
Report Number3002808148-2024-32352
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K001203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-G20-29R
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-