Catalog Number 8065752914
Device Problem Overheating of Device (1437)
Patient Problem Eye Burn (2523)
Event Date 03/22/2024
Event Type  malfunction  
Event Description
A customer reported that during cataract surgery, an ophthalmic handpiece becomes hot and patient experienced thermal burn.The surgery was completed after replacing the handpiece with another one.Patient details have not been provided.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.The device was returned for investigation.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The device was received for investigation.A visual assessment of the returned sample revealed no obvious nonconformity.The device was connected to a calibrated resistance test box which found the input and output impedance, resistance, and crystal dissipation tests were within specification.A flow rate test was performed on the irrigation and aspiration lines of the device which found the device to meet specifications.The returned sample was connected to a calibrated system.The device tuned successfully and completed a five-minute burn-in test with the system set at 100% ultrasonic and torsional power.During the burn-in test, the temperature of the device was measured specifications.The device was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements which found the device to meet specifications.The device was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, ACTIVE SENTRY HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key19149766
MDR Text Key340718096
Report Number2028159-2024-00583
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657529148
UDI-Public(01)00380657529148(17)330317(11)230320(10)15EVA3(21)2301490809X
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K161794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752914
Device Lot Number15EVA3
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/24/2024
Initial Date FDA Received04/20/2024
Supplement Dates Manufacturer Received07/10/2024
Supplement Dates FDA Received08/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM (ACTIVE SENTRY).
-
-