Model Number CLV-190
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/26/2024
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to olympus for evaluation of the reported issue.In speaking with olympus technical assistance center (tac), the customer reported that the xenon light source displayed error code b30 (scope communication error).Troubleshooting efforts were made and the customer was instructed to follow the quick reference guide on the b30 (scope communication error).It was advised to power off both the cv-190 and the clv-190, remove the scope, clean the contact points on the scope with an alcohol prep pad, wait for the scope to dry, reconnect the scope, power everything back up, and observe the result.Additionally, the customer was also instructed to always disconnect or connect the scope when light source is powered off.If the b30 error code (scope communication error) or e216 error code (scope communication error) comes up.Both the cv-190 and the clv-190 must be powered off.The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the xenon light source displayed error code b30 (scope communication error).The issue occurred during preparation for use.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19150104
MDR Text Key341473803
Report Number3002808148-2024-03702
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/12/2024
Initial Date FDA Received04/21/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CV-190-EVIS EXERA III VIDEO SYSTEM CENTER.
-
-