• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY RF DR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number PM2240
Device Problem Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/02/2024
Event Type  malfunction  
Event Description
During the generator changes procedure, it was noted the right ventricular (rv) lead was failed to remove form the pacemaker header and the rv lead was damaged.The pacemaker was explanted and replaced.Meanwhile the same rv lead was used.The patient was in stable condition throughout the procedure.
 
Manufacturer Narrative
The reported event of could not remove the ventricular lead from the header was confirmed.Analysis revealed this device has the original scalable brady pacemaker (sbp) header design, which is known to have lead insertion and removal difficulties.This caused the v-lead to become stuck and un-removeable from the header.This anomaly with the header caused the v-lead was pulled apart by the physician while trying to remove it out of the connector.This shows the high abnormal force needed to remove leads from the connectors of the original sbp header design.Actions have already been taken to prevent reoccurrence.A header re-design has been implemented to correct this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19150134
MDR Text Key340809954
Report Number2017865-2024-39216
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507073
UDI-Public(01)05414734507073(10)4657419(17)151231
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2015
Device Model NumberPM2240
Device Lot Number4657419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/21/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL STS LEAD.
Patient SexFemale
-
-