• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. XPS® BUR - UNKNOWN; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. XPS® BUR - UNKNOWN; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number XOM UNK DRILLBUR
Device Problems Break (1069); Fitting Problem (2183); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/31/2024
Event Type  malfunction  
Manufacturer Narrative
H3: visually, the bur tang was dislodged from the proximal end of a curved bur when returned.The length of the bur tang shall be 0.750 ± 0.005 inches and the actual measurement was 0.748 inches.The outside diameter of the distal end of the bur tang shall be 0.0919 ± 0.000 inches and the actual measurement was 0.0921 inches.The outside diameter of the proximal end of the bur tang shall be 0.056 ± 0.002 inches and the actual measurement was 0.056 inches.All the dimensional measurements were in specification.Functional testing could not be performed due to the broken state of the device.In the returned condition, there was an out of specification condition that was related to the complaint (due to physical damage).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
A health care professional reported that prior to use, bur was not seating flush in visao handpiece.There was no patient impact.Analysis of handpiece found that there was a broken bur stuck in handpiece.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® BUR - UNKNOWN
Type of Device
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key19150777
MDR Text Key340680420
Report Number1045254-2024-00587
Device Sequence Number1
Product Code EQJ
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberXOM UNK DRILLBUR
Device Catalogue NumberXOM UNK DRILLBUR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/26/2024
Initial Date FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-