H3: visually, the bur tang was dislodged from the proximal end of a curved bur when returned.The length of the bur tang shall be 0.750 ± 0.005 inches and the actual measurement was 0.748 inches.The outside diameter of the distal end of the bur tang shall be 0.0919 ± 0.000 inches and the actual measurement was 0.0921 inches.The outside diameter of the proximal end of the bur tang shall be 0.056 ± 0.002 inches and the actual measurement was 0.056 inches.All the dimensional measurements were in specification.Functional testing could not be performed due to the broken state of the device.In the returned condition, there was an out of specification condition that was related to the complaint (due to physical damage).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|