Model Number 5076-52
Device Problems Pacing Problem (1439); Device Sensing Problem (2917); Device Dislodged or Dislocated (2923)
Patient Problems Arrhythmia (1721); Bradycardia (1751); Discomfort (2330)
Event Date 04/09/2024
Event Type  Injury  
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that the patient experienced palpitations and chest discomfort.Arrhythmia with premature atrial contraction and premature ventricular contractions (pvc) were also noted.The right atrial (ra) lead exhibited increase in thresholds.The patient's pulse rate was at thirties to forty beats per minute.Hospitalization was required as a result of the event.The ra and right ventricular (rv) lead exhibited sensing and pacing failure.Ra and rv lead dislodgement was confirmed.The leads were reprogrammed and repositioned.The leads remain in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS LEAD MRI SURESCAN
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN  486056
Manufacturer (Section G)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN   486056
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key19151102
MDR Text Key340676117
Report Number3008973940-2024-02860
Device Sequence Number1
Product Code DTB
UDI-Device Identifier00681490124812
UDI-Public00681490124812
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
P930039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number5076-52
Device Catalogue Number5076-52
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/16/2024
Initial Date FDA Received04/22/2024
Date Device Manufactured10/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
W2DR01 IPG, 5076-45 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
-
-