Model Number U225
Device Problems Over-Sensing (1438); Under-Sensing (1661)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2024
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) recorded an episode of atrial tachy response (atr) events.Upon review, technical services (ts) determined that an atr had occurred due to far-field oversensing in the right atrial channel.It was also noted that the far-field oversensing caused functional undersensing of some atrial beats.Ts suggested adjusting the sensing threshold of the atrial channel.The sensing threshold was adjusted and the oversensing was resolved.This device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST CRT-P
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key19151602
MDR Text Key340926893
Report Number2124215-2024-24177
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559426
UDI-Public00802526559426
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/09/2023
Device Model NumberU225
Device Catalogue NumberU225
Device Lot Number718967
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2024
Initial Date FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
-
-