• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART MRX ALS MONITOR; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART MRX ALS MONITOR; DEFIBRILLATOR Back to Search Results
Model Number M3536A
Device Problem Failure to Discharge (1169)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2024
Event Type  malfunction  
Event Description
This report is based on information provided by philips authorized service provider (asp) and with an investigation documented by philips complaint handling.Philips received a complaint on the heartstart mrx defibrillator indicating that the device cannot be discharged sometimes.The device was evaluated onsite by philips asp with a check of the error logs.Based on the results of the analysis, it was determined that the product has malfunctioned and cause of the reported problem was due to a defective high voltage capacitor.The issue was resolved after the defective high voltage capacitor was replaced and the product is back in service.Based on the information available and results of additional analysis, no further action is necessary at this time.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.If additional information is received, the complaint file will be reopened for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX ALS MONITOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
tanya deschmidt
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key19151647
MDR Text Key341372929
Report Number3030677-2024-01463
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838006652
UDI-Public00884838006652
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Device Catalogue NumberM3536A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/28/2024
Initial Date FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-