The guide wire was returned for analysis.The guide wire was fractured at the proximal spring tip.Scanning electron microscopy (sem) was performed on the fractured proximal spring tip coil and identified evidence of rotational damage.This evidence suggests the spinning oad driveshaft contacted the guide wire spring tip, resulting in the guide wire fracture.The root cause of the fracture is considered use not consistent with the instructions for use (ifu).The ifu cautions to maintain at least 5 mm between the proximal end of the viperwire guide wire spring tip and the oad drive shaft tip to prevent contact of the drive shaft tip with the guide wire spring tip.Further advance the viperwire guide wire, as necessary, to maintain the 5 mm minimum distance.The material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|
The diamondback 360 coronary orbital atherectomy device (oad) was used to treat a heavily calcified left anterior descending artery (lad).Following atherectomy treatment, when the oad and viperwire advance guide wire were removed, the viperwire spring tip was observed fractured.A snare was used to successfully retrieve the component.The patient was stable.The physician was unsure when the fracture occurred as there were no issues experienced during the procedure.
|