CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE); CORONARY ATHERECTOMY DEVICE (GUIDE WIRE)
|
Back to Search Results |
|
Model Number GWC-12325LG-FT |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/25/2024 |
Event Type
malfunction
|
Event Description
|
The target treatment area was a (b)(4) stenosed, moderately calcified and moderately tortuous left anterior descending artery (lad).The diamondback 360 coronary orbital atherectomy device (oad) was spun three times on low speed for treatment.The oad was advanced again while on glideassist and the viperwire advance guide wire became unstable and the oad contacted the viperwire spring tip, resulting in loss control of the viperwire.The oad and viperwire were removed together.Ex vivo, the viperwire was observed fractured.There were no components left in vivo.Angioplasty was performed to complete the procedure.The patient was stable and did not experience impact from the reported issue.
|
|
Manufacturer Narrative
|
The guide wire was returned for analysis.The guide wire was fractured at the spring tip solder bond.Scanning electron microscopy (sem) was performed on the guide wire core and spring tip sections, sem identified evidence of rotational damage on the spring coil and ductile torsion on the core wire fracture face.This evidence is consistence with the spinning driveshaft contacting the spring tip leading to the fracture.The root cause of the failure is considered use not consistent with the instruction of use (ifu).The ifu warns to use fluoroscopy to monitor movement of the radiopaque diamondback 360 coronary orbital atherectomy device (oad) shaft tip in relation to the radiopaque viperwire guide wire spring tip.Always keep more than 5mm of spacing between the distal end of the oad driveshaft and the proximal end of the guide wire spring tip.[if the distance between the shaft tip and the viperwire guide wire spring tip is insufficient, the shaft tip may contact the guide wire spring tip and result in dislodging the guide wire spring tip.] the material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
|
|
Search Alerts/Recalls
|
|
|