• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2 Back to Search Results
Model Number 863283
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2024
Event Type  malfunction  
Manufacturer Narrative
A philips biomedical equipment technician went to the customer site to evaluate the device and identified that power cable was pulled halfway away from the vs4 device, possibly causing it to short.The biomed resecured the connection before plugging the unit back.The biomed could not duplicate the reported problem.Based on the information available and the testing conducted, the cause of the reported problem was user error and not a device malfunction.The reported problem was not confirmed.No part was replaced.The investigation concludes that no further action is required at this time.Reporting institution phone: n/a reporter phone: n/a.
 
Event Description
It was reported the vitals machine emitted sparks from the outlet.The device was not in use on a patient at the time of event, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESIGNS VS4 NBP, SPO2
Type of Device
SURESIGNS VS4 NBP, SPO2
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key19154278
MDR Text Key340731303
Report Number1218950-2024-00285
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838087095
UDI-Public00884838087095
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number863283
Device Catalogue Number863283
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/01/2024
Initial Date FDA Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-