Model Number CD1411-36Q
Device Problems Failure to Interrogate (1332); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that patient presented in clinic for routine follow-up.Upon interrogation, it was found that patient's implantable cardioverter defibrillator (icd) was in backup mode due to firmware update related reset.It was further noted that device was not able to interrogate.The device was able to reset to normal setting by programming intervention.The patient status was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE VR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key19159985
MDR Text Key340814052
Report Number2017865-2024-39312
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734507738
UDI-Public(01)05414734507738(10)A000020742(17)180430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2018
Device Model NumberCD1411-36Q
Device Lot NumberA000020742
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/03/2024
Initial Date FDA Received04/22/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA.
Patient Age79 YR
Patient SexMale
-
-