A.1.-a.5.There was no patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in milan, italy.Through follow-up communication with the customer, livanova learned that the device was cleaned regularly as per instructions for use, disposable gloves are used solely for heater cooler 3t device during cleaning, patient reusable blankets are not used by the hospital, water quality monitoring is applied and the h2o2 checked daily.In addition, when the device is not used it is it stored full of water and h2o2 is checked daily even during days of non-use.H2o2 is added when the water in the tanks is changed (each 7 days), a disposable pall-aquasafe water filter with 0.2m membrane or equivalent performance filter is used for tap water, prior to initial operation and prior to storing the heater-cooler the surfaces and water circuits are disinfected, 3t aerosol collection set is replaced after allowed use period (7 days) and blue tubing set used with the device is replaced each year.The device is placed inside the operation theater during use with the fan of the device positioned opposite to patient, at an estimated distance between the surgery field and the device of two (2) meters.Despite no evident or systematic deviation from device instruction for use could be identified in this specific case, however, the source of contamination is most likely related to location where device is used/stored.
|