SMITH & NEPHEW, INC. STEM IMPACTOR ROD; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number 71364011
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2024
Event Type  malfunction  
Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event Description
It was reported that during thr surgery, the tip of the stem impactor rod broke off when the physician impacted the stem.The tip remained lodged in the stem and was retrieved by pulling it out with pliers.There was only about a 3 minutes delay in the case.X-ray was not needed.The procedure was completed using a s+n back-up device.No injury was reported as a consequence of this issue.
 
Manufacturer Narrative
The associated device was returned and evaluated.The visual inspection revealed that the tip of the device broke off.The tip of the device was not returned and the device has signs of wear, dents, and scratches.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of previously escalated cases concluded that this failure was further investigated and addressed through design change evaluations.The batch number under this complaint was manufactured before these actions were initiated.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for additional corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM IMPACTOR ROD
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19164413
MDR Text Key341055844
Report Number1020279-2024-00808
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010199270
UDI-Public03596010199270
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/18/2010
Device Catalogue Number71364011
Device Lot Number01110850
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/03/2024
Initial Date FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-