Model Number V60 V60PLUS VENTILATOR
Device Problem Display Difficult to Read (1181)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/16/2024
Event Type  malfunction  
Manufacturer Narrative
G1: contact office phone: (b)(6).
 
Event Description
Philips received a complaint by the customer on the v60 indicating that the backlight was going out on the display and the display is dim.It was reported there was no patient involvement at the time the issue was discovered.It was reported that the backlight was going out on the display and the display was dim.The customer requested part number to order and replace the light crystal display (lcd).The remote service engineer (rse) explained to the customer an upgrade of the lcd from 1st generation to 2nd generation was needed.The rse provided the customer with the part number and pricing of the lcd.Device evaluation and repair are pending.
 
Manufacturer Narrative
H10: the customer replaced the light crystal display (lcd) to resolve the reported issue.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.Although requested, the defective parts were not received at this time.A supplemental report will be submitted if the part is received and evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19164474
MDR Text Key340942412
Report Number2518422-2024-21633
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/16/2024
Initial Date FDA Received04/23/2024
Supplement Dates Manufacturer Received04/29/2024
Supplement Dates FDA Received04/30/2024
Date Device Manufactured05/03/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-