It was reported that spline bunching was experienced along with forward facing splines, and the guidewire was stuck.During a pulse field ablation (pfa) procedure to treat paroxysmal atrial fibrillation (paf), a farawave catheter was selected for use.The catheter was flushed and prepped per the instructions for use (ifu).During deployment in the right superior pulmonary vein, spline bunching was noted.The catheter was rotated in an attempt to correct the issue, but splines remained bunched.The device was withdrawn into the sheath and removed from the patient anatomy.The sheath was angled during retraction into the sheath.The splines were observed facing forward.It was noted that the guidewire was stuck; no tissue was seen at the tip of the catheter or guidewire.The guidewire was extended beyond the catheter tip.The catheter was not deployed without a guidewire inserted; a non-boston scientific (bsc).035 180cm 1.5mm j-tip guidewire was used.The guidewire had not been reloaded into the catheter.There were no issues with flushing the catheter.The catheter was replaced, and the procedure was completed successfully without patient complications.The device is expected to be returned for analysis.
|