H4: the lot was manufactured between january 5, 2024 and january 7, 2024.H11: the device was received for evaluation.A visual inspection was performed which observed the fill port tube was not sealed into the bag.The port was found seated outside the port area which would allow for a leak.The reported condition was verified.The cause of the condition is most likely due to the weld sealing process during manufacturing.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|