• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX800 PATIENT MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX800 PATIENT MONITOR Back to Search Results
Model Number 865240
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/17/2024
Event Type  malfunction  
Manufacturer Narrative
The following functional tests and communications were performed.A philips remote service engineer (rse) spoke to the customer and confirmed the issue that speaker malfunction inop was displayed and no sound was producing from the device.Based on the information available and the testing conducted, the cause of the reported problem was a faulty mainboard.The reported problem was confirmed.Nemo (non-engineering material only) service was agreed upon.The customer ordered a replacement mainboard to resolve the issue.The customer was provided a replacement mainboard to resolve the issue.It has been concluded that no further action is required at this time.
 
Event Description
The customer reported there was a speaker malfunction inoperative error (inop) message on the monitor, and there was no sound.The device was not in clinical use at the time the issue was discovered.No adverse event or patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX800 PATIENT MONITOR
Type of Device
INTELLIVUE MX800 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key19166832
MDR Text Key341432934
Report Number9610816-2024-00222
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838020733
UDI-Public00884838020733
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865240
Device Catalogue Number865240
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2024
Initial Date FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-