E4: medwatch (b)(4).Philips became aware of the issue on february 22, 2024.A field safety notification / urgent medical device correction is being sent to all of the affected customers.It is recommended that ups devices intended for operation in network and server rooms be kept in a temperature and humidity controlled environment, ensuring adequate airflow around the ups to prevent ups device failure.A field change order (fco) is in the process of being released.A philips representative will contact affected customers to schedule a visit from a philips field service engineer who will replace the malfunctioning ups.A philips representative will contact affected customers to schedule a visit from a philips field service engineer who will replace the malfunctioning ups.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|