Model Number CLV-190 |
Device Problem
Intermittent Loss of Power (4016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/16/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was intermittent power, while using the lightsource.There was no patient harm associated with the event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation. a review of the device history record found no deviations, that could have caused or contributed to the reported issue.It has been over 6 years, since the subject device was manufactured.Based on the results of the investigation.A definitive root cause could not be determined, since the device was not returned for device evaluation.Olympus will continue to monitor field performance for this device. .
|
|
Event Description
|
It was reported, that the light source was shutting off.The issue occurred, during a procedure.There was no patient harm associated with the event.Additional information received, from the customer.Reported, that they discovered there was a loose wire.And the device was now working.
|
|