• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VULCAN CUSTOM DENTAL VULCAN CUSTOM DENTAL SURGICAL GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VULCAN CUSTOM DENTAL VULCAN CUSTOM DENTAL SURGICAL GUIDE Back to Search Results
Model Number PRINT003
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/13/2024
Event Type  malfunction  
Manufacturer Narrative
A review of the drilling protocol and surgical report indicate the case was properly planned and no defects were noted on the returned surgical guide.In addition, a radiograph taken after the implant was placed, was was sent to dr.Cleber silva for review and per dr.Silva, the radiograph sent to him by the doctor shows the implant appears to have been placed as planned, he also stated that "immediate implants usually go at an angle to engage native bone but this implant was straight as it had at least 5.0 mm of post-apical bone engagement.".
 
Event Description
Utilizing dental surgical guide print003, clinician states that they could not initially get the implant to engage with bone and after it was engaged, the clinician felt the placement was off, so implant was removed, patient was grafted and sent home.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VULCAN CUSTOM DENTAL SURGICAL GUIDE
Type of Device
DENTAL SURGICAL GUIDE
Manufacturer (Section D)
VULCAN CUSTOM DENTAL
2300 riverchase center
suite 825
birmingham AL 35244
Manufacturer (Section G)
VULCAN CUSTOM DENTAL
2300 riverchase center
suite 825
birmingham AL 35244
Manufacturer Contact
brett muetzel
2300 riverchase center
suite 825
birmingham, AL 35244
8444842301
MDR Report Key19169135
MDR Text Key341417166
Report Number3012481042-2024-00032
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Dentist
Type of Report Initial
Report Date 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPRINT003
Device Catalogue NumberPRINT003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age30 YR
Patient SexPrefer Not To Disclose
-
-