Model Number PRINT003 |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/13/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
A review of the drilling protocol and surgical report indicate the case was properly planned and no defects were noted on the returned surgical guide.In addition, a radiograph taken after the implant was placed, was was sent to dr.Cleber silva for review and per dr.Silva, the radiograph sent to him by the doctor shows the implant appears to have been placed as planned, he also stated that "immediate implants usually go at an angle to engage native bone but this implant was straight as it had at least 5.0 mm of post-apical bone engagement.".
|
|
Event Description
|
Utilizing dental surgical guide print003, clinician states that they could not initially get the implant to engage with bone and after it was engaged, the clinician felt the placement was off, so implant was removed, patient was grafted and sent home.
|
|