• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VULCAN CUSTOM DENTAL VULCAN CUSTOM DENTAL SURGICAL GUIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VULCAN CUSTOM DENTAL VULCAN CUSTOM DENTAL SURGICAL GUIDE Back to Search Results
Model Number PRINT003
Device Problem Use of Device Problem (1670)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2024
Event Type  malfunction  
Manufacturer Narrative
A review of the drilling protocol and surgical report indicate the case was properly planned and no defects were noted on the returned surgical guide.Dr.Cleber silva reviewed the case and stated this was a case with severe peridontal disease and a lot of missing bone which should have warranted immediate placement.The scan the doctor sent in was 3 months old, meaning that in addition to bone already lost at the time of scanning, a site with periodontal disease can undergo rapid and accelerated bone loss in 3 months, particularly if localized infections occur, which is quite common.Dr.Silva's professional opinion is either (1) more bone was lost in the 3 months between scanning and surgery or (2) bone was lost during extraction.The surgical guide was not the cause of the problem.
 
Event Description
Utilizing surgical guide print003, clinician placed an implant in location #26 and after placing realized there was no buccal bone on the implant.The plan had plenty of buccal bone and no bone came out with the extraction.Clinician ended up removing the implant and placing another one freehand.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VULCAN CUSTOM DENTAL SURGICAL GUIDE
Type of Device
DENTAL SURGICAL GUIDE
Manufacturer (Section D)
VULCAN CUSTOM DENTAL
2300 riverchase center
suite 825
birmingham AL 35244
Manufacturer (Section G)
VULCAN CUSTOM DENTAL
2300 riverchase center
suite 825
birmingham AL 35244
Manufacturer Contact
brett muetzel
2300 riverchase center
suite 825
birmingham, AL 35244
8444842301
MDR Report Key19169271
MDR Text Key341675185
Report Number3012481042-2024-00033
Device Sequence Number1
Product Code NDP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Dentist
Type of Report Initial
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPRINT003
Device Catalogue NumberPRINT003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2024
Initial Date FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/03/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age30 YR
Patient SexPrefer Not To Disclose
-
-