RESPIRONICS, INC. BIPAP A40 SYSTEM SILVER; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number R1111177
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2024
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator inoperative condition occurred on a bipap a40 device.There was no harm or injury reported.The device was returned to the manufacturer's service center for evaluation.During the evaluation of the device, the customers complaint was not confirmed, but an event log error was observed.The device's main board was replaced to address the issue.
 
Manufacturer Narrative
H3 other text : device not returned at manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40 SYSTEM SILVER
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key19169459
MDR Text Key340903013
Report Number2518422-2024-21867
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121623
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberR1111177
Device Catalogue NumberR1111177
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/10/2024
Initial Date FDA Received04/23/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-