Model Number 865350 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/02/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information regarding the reported event and the investigation is ongoing.A follow-up report will be submitted upon completion of the investigation.
|
|
Event Description
|
During evaluation at bench repair, it was identified that the mx40 had no audio.The device was not in use at time of event, there was no adverse event reported.
|
|
Manufacturer Narrative
|
Diagnostic/functional testing was performed at the philips authorized repair facility.The bench indicated that speaker had no sound in the mt56060 tool.The speaker was confirmed to be defective.The speaker was replaced by the bench technician.The device was operational after repair of the speaker and was returned to the customer.
|
|