MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX450 PATIENT MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX450 PATIENT MONITOR Back to Search Results
Model Number 866062
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/16/2024
Event Type  malfunction  
Manufacturer Narrative
Section e reporting institution phone #: (b)(6).A philips field service engineer (fse) visited onsite and unable to confirm the reported issue ¿speaker error and no sound".The fse turned on the monitor and no error message appeared on the device and the fse confirmed that alarms are audible.The fse replaced the loudspeaker assembly in case the problem is intermittent.
 
Event Description
It was reported intellivue mx450 patient monitor displays a speaker malfunction error and no sound is coming from the device.The device was in use on a patient.There was no report of patient or user harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX450 PATIENT MONITOR
Type of Device
INTELLIVUE MX450 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key19172179
MDR Text Key341201456
Report Number9610816-2024-00224
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838038769
UDI-Public00884838038769
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K130849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866062
Device Catalogue Number866062
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/16/2024
Initial Date FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-