RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/08/2024
Event Type  malfunction  
Event Description
Philips received a complaint indicating that the v60 ventilator had multiple error codes that resulted from a check vent: ovp circuit failure - error code 1127.The other error codes were 111b (35 v supply failed), 111d (mc pcba adc failed), 1117 (3.3 v supply failed), and 1118 (5 v supply failed.).It is unknown if the device was in use on a patient at the time of the event.There was no patient or user harm reported.
 
Manufacturer Narrative
Insufficient information is available to determine the resolution of the event.The customer was provided with a quote to have the device evaluated and repaired; however, the customer didn¿t accept the quote and no work order was generated.It is unknown what the outcome of the service event was, and no further information was obtained.The investigation concludes that no further action is required at this time.If the decision is made to have the device evaluated and repaired a new service order will be opened and will be captured through philip's normal complaint procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19172349
MDR Text Key340915382
Report Number2518422-2024-21878
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/08/2024
Initial Date FDA Received04/24/2024
Supplement Dates Manufacturer Received04/18/2024
Supplement Dates FDA Received04/25/2024
Date Device Manufactured06/14/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-