RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Complete Loss of Power (4015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/08/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint indicating that the v60 ventilator had multiple error codes that resulted from a check vent: ovp circuit failure - error code 1127.The other error codes were 111b (35 v supply failed), 111d (mc pcba adc failed), 1117 (3.3 v supply failed), and 1118 (5 v supply failed.).It is unknown if the device was in use on a patient at the time of the event.There was no patient or user harm reported.
|
|
Manufacturer Narrative
|
Insufficient information is available to determine the resolution of the event.The customer was provided with a quote to have the device evaluated and repaired; however, the customer didn¿t accept the quote and no work order was generated.It is unknown what the outcome of the service event was, and no further information was obtained.The investigation concludes that no further action is required at this time.If the decision is made to have the device evaluated and repaired a new service order will be opened and will be captured through philip's normal complaint procedure.
|
|