• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE® CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE® CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF283
Device Problems Material Integrity Problem (2978); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2024
Event Type  malfunction  
Event Description
It was reported that during a cryo ablation procedure, while ablating the right lower pulmonary vein (pv), a system notice was received indicating that the safety system detected a high level of refrigerant flow and stopped the injection.The freeze stopped and the balloon catheter could not be reinflated and a system notice was received indicating that the safety system detected a compromised outer vacuum.The coaxial umbilical cable was replaced without resolve.The balloon catheter was replaced to resolve the issue.Upon removal it was found that the inner and outer balloon layers were separated. the case was completed with cryo. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was reported that during a cryo ablation procedure, while ablating the right lower pulmonary vein (pv), a system notice was received indicating that the safety system detected a high level of refrigerant flow and stopped the injection.The freeze stopped and the balloon catheter could not be reinflated and a system notice was received indicating that the safety system detected a compromised outer vacuum.The coaxial umbilical cable was replaced without resolve.The balloon catheter was replaced to resolve the issue.Upon removal it was found that the inner and outer balloon layers were separated. the case was completed with cryo. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the 2af283 balloon catheter with lot number 25266 and the data files were returned and analyzed.The patient file showed 10 applications were performed using the faulty balloon catheter, and 3 applications were performed using the replacement balloon catheter.The patient file also showed system notice #50032 (the safety system has detected a compromised outer vacuum) during inflation/transition at application #6 to 10 with the first catheter.The patient file for the second catheter showed system notice #50001 (the vacuum is disabled due to a problem with the coaxial umbilical cable) during ablation at application #11 and 13; and system notice 50030 (the safety system has detected a high level of refrigerant flow and stopped the injection) during transition at application #12.No anomalies were identified during external visual inspection of the balloon, shaft, and handle segments.The catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for 12 applications on the reported event date.During functional testing, the console terminated the application and triggered system notice #50032.During pressure testing of the balloon segment, an inner balloon breach was observed.Dissection of the balloon segment identified an inner balloon material fracture.In conclusion, the reported 'double balloon damage" was confirmed during analysis, and the balloon catheter failed the returned product inspection due to an inner balloon breach.The breach was a material/crack on the surface of the inner balloon.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE® CARDIAC CRYOABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19173287
MDR Text Key341430689
Report Number3002648230-2024-00168
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P100010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2AF283
Device Catalogue Number2AF283
Device Lot Number25266
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/22/2024
Initial Date FDA Received04/24/2024
Supplement Dates Manufacturer Received05/01/2024
Supplement Dates FDA Received05/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
Patient Weight80 KG
-
-