Catalog Number 71368569
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2024
Event Type  malfunction  
Event Description
It was reported that, during thr surgery, the piece that hold the cup of the r3 offset impactor was broken off.Nothing fell into the patient.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported as a consequence of this issue.
 
Manufacturer Narrative
Internal reference number: (b)(4).H10: smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
 
Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.The visual inspection revealed that the impactor tip was not returned and a device breakage cannot be confirmed.In addition, no breakage in the returned components could be identified.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that a similar event was previously identified, and prior actions were performed.It has been concluded that there is a potential for breakage if used after the expected lifetime or excessive force is used as this device is a reusable instrument and it is expected to wear over time.Also, the failure rate of this issue is within expected and documented in the risk file.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, actions were previously opened to address this issue.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 OFFSET IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19175180
MDR Text Key341056562
Report Number1020279-2024-00818
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010600240
UDI-Public03596010600240
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71368569
Device Lot Number22BM19937
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/03/2024
Initial Date FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/25/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-