H10: h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A clinical review states that based on the limited information provided the clinical root cause of the reported events could not be determined and we are currently unable to rule out a procedural variance as a contributing factor to the reported event, which does not represent a device malfunction.The device ifu does caution that ¿use of excessive force during insertion can cause failure of the suture anchor or insertion device.¿ the osteoraptor anchor is implantable, biocompatibility is not an issue.If fragments from the broken osteoraptor anchor were retained, local irritation/discomfort, and/or migration of the anchors should not be ruled out.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
|