• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HEAD 45MM; HIP INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US QUICKSET ACE GRATER HEAD 45MM; HIP INSTRUMENTS : REAMERS Back to Search Results
Catalog Number 244000545
Device Problem No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2024
Event Type  malfunction  
Event Description
Instruments were reported for unknown reason.The event did not happen in surgery.Additional event information: visual analysis of the device found rounding of cutting edges with signs of worn out at the surface and corrosion on both inside and outside of the quickset ace grater head.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : according to the information received, instruments were reported for unknown reason.The event did not happen in surgery.The product was returned to depuy synthese for evaluation.The depuy synthese team conducted a visual inspection of the returned device.Visual analysis of the device found rounding of cutting edges with signs of worn out at the surface and corrosion on both inside and outside of the quickset ace grater head.The device exhibits damage consistent with repeated use.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Where corrosion/oxidation was identified around the cutting surface of the device, alongside damage consistent with normal wear, this suggests that the passive layer of the metal has been worn down from repeated use (due to normal use damage) allowing for the metallic surface underneath to become susceptible to corrosion/oxidation.The lifecycle requirements of the device are event related and depend on the use in clinical practice, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.The overall complaint was confirmed as the observed condition of the quickset ace grater head would have contributed to the device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.As part of depuy synthese quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese, or its employees that the report constitutes an admission that the product, depuy synthese, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET ACE GRATER HEAD 45MM
Type of Device
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19175856
MDR Text Key341241822
Report Number1818910-2024-09006
Device Sequence Number1
Product Code HTO
UDI-Device Identifier10603295123927
UDI-Public10603295123927
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number244000545
Device Lot NumberSO2054551
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/23/2024
Initial Date FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-