It was reported that the patient underwent an atrial fibrillation ablation procedure with a qdot micro¿ catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the proceudre, a temperature slope is too high message was displayed on the ngen remote when coming on ablation and ablation would be cut off.The cable was replaced with no resolution.When the catheter was replaced, the issue was resolved.No patient consequences were reported.
|
The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.Visual inspection, temperature and impedance test, and patency test of the returned device were performed following bwi procedures.Visual analysis revealed reddish material and a hole in the surface of the pebax.The temperature and impedance test was performed and the device was found working correctly.No temperature or impedance issues were observed.A patency test was performed, and the device was flushing correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device [31231689l] number, and no internal actions related to the reported complaint condition were identified.The reddish material inside the pebax could be related to the temperature issue reported by the customer; therefore, the customer complaint was confirmed.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instruction for use states: when cleaning the tip electrode, be careful not to twist the tip electrode with respect to the catheter shaft; twisting may damage the tip electrode bond and loosen the tip electrode, or may damage the contact force sensor.In addition, in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle.If the radiofrequency (rf) generator does not display temperature, verify that the appropriate cable is plugged into the rf generator.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|