MAUDE Adverse Event Report: BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC QDOT MICRO¿ CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Catalog Number D139505
Device Problem Insufficient Cooling (1130)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2024
Event Type  malfunction  
Event Description
It was reported that the patient underwent an atrial fibrillation ablation procedure with a qdot micro¿ catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the proceudre, a temperature slope is too high message was displayed on the ngen remote when coming on ablation and ablation would be cut off.The cable was replaced with no resolution.When the catheter was replaced, the issue was resolved.No patient consequences were reported.
 
Manufacturer Narrative
The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.Visual inspection, temperature and impedance test, and patency test of the returned device were performed following bwi procedures.Visual analysis revealed reddish material and a hole in the surface of the pebax.The temperature and impedance test was performed and the device was found working correctly.No temperature or impedance issues were observed.A patency test was performed, and the device was flushing correctly, no obstructed holes were observed.No irrigation issues were observed.A manufacturing record evaluation was performed for the finished device [31231689l] number, and no internal actions related to the reported complaint condition were identified.The reddish material inside the pebax could be related to the temperature issue reported by the customer; therefore, the customer complaint was confirmed.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.The instruction for use states: when cleaning the tip electrode, be careful not to twist the tip electrode with respect to the catheter shaft; twisting may damage the tip electrode bond and loosen the tip electrode, or may damage the contact force sensor.In addition, in order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle.If the radiofrequency (rf) generator does not display temperature, verify that the appropriate cable is plugged into the rf generator.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QDOT MICRO¿ CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key19176527
MDR Text Key341017654
Report Number2029046-2024-01356
Device Sequence Number1
Product Code OAE
UDI-Device Identifier10846835016758
UDI-Public10846835016758
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P210027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 04/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD139505
Device Lot Number31231689L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2024
Initial Date Manufacturer Received 04/01/2024
Initial Date FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NGEN GENERATOR
-
-