• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4.5 HEALIX BR ANCHOR W/OCORD; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4.5 HEALIX BR ANCHOR W/OCORD; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 222229
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2024
Event Type  malfunction  
Event Description
This is report 2 of 2 for (b)(4).It was reported from china that during a rotator cuff repair procedure, it was observed that the anchor on the 4.5 healix br anchor w/ocord device broke.Changed to another one to continue the procedure but the same problem happened again.Another like device was used to complete the procedure.There were no adverse consequences to the patient.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: (b)(4) investigation summary the product was not returned, however a photo was provided for review.The photo investigation revealed that 4.5 healix br anchor w/ocord was missing a broken part from the distal end to the middle of the anchor.No anomalies could be observed on the inserter.The overall complaint was confirmed as the observed condition of the 4.5 healix br anchor w/ocord would contribute to the complained device issue.Based on the investigation findings, the potential cause for the broken anchor can be traced to off axis insertion and levering during insertion.As per ifu: inserting the awl less than the specified depth, axial misalignment or levering with the anchor upon insertion, may result in anchor fracture, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review a review of the batch manufacturing records for finished lot 226l798 was conducted, and no related non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5 HEALIX BR ANCHOR W/OCORD
Type of Device
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
MEDOS INT MITEK
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key19176730
MDR Text Key341019046
Report Number1221934-2024-01333
Device Sequence Number1
Product Code MAI
UDI-Device Identifier10886705007769
UDI-Public10886705007769
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K150209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number222229
Device Lot Number226L798
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/09/2024
Initial Date FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-