The customer reported the autopulse platform (b)(6) was brought in with blood in the head restraint wires that cannot be cleaned out.The head restraint wires are broken and frayed and need to be replaced.Also, the display screen on the board is dark and hard to read.However, the customer reported the board works fine.No patient involvement.
|
The reported complaint that the autopulse platform (b)(6) was brought in with blood in the head restraint wires and the head restraint wires are broken and frayed was confirmed during the visual inspection.Both wire strands of the top cover were heavily frayed.The damaged head restraints do not render the autopulse platform non-functional.The likely root cause for the reported complaint was due to mishandling, however contribution from wear and tear cannot be ruled out due to the age of the device.The autopulse platform was manufactured in 2007 and is over 17 years old, well past its expected serviceable life of 5 years.The top cover will be replaced to address the reported complaint.The secondary complaint that the display screen on the autopulse platform (b)(6) is dark and hard to read was confirmed during the visual inspection.The root cause for the reported complaint was a failed lcd, likely attributed to the age of the device.The lcd will be replaced to remedy the issue.Upon further visual inspection, the bottom and front covers had multiple cracks in the screw well area and the load plate cover has noticeable cuts, unrelated to the reported complaint.The observed physical damages appeared to be the characteristics of user mishandling.The front, bottom, and load plate covers will be replaced to address the issue.Upon review of the archive data, no significant discrepancies were found.The autopulse platform passed the initial functional testing without any fault or error.Upon service completion, the platform will be subjected to final functional testing to ensure that the device passes all testing criteria.Historical complaints were reviewed for service information related to the reported complaint, and there were no similar complaints reported for autopulse platform with sn (b)(6).
|
Service was not performed on the autopulse platform with serial number (b)(6).Due to the age of the device, and zoll proposed a newer platform as a replacement.The customer accepted the offer and their original device (sn (b)(6)) was scrapped at zoll.The new ap platform with serial number (b)(6), which was manufactured on june 22, 2022, was shipped to the customer.
|