MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number UDSX500S11F
Device Problems Unexpected Shutdown (4019); Excessive Heating (4030)
Patient Problems Sleep Dysfunction (2517); Dry Mouth (4485)
Event Date 01/31/2024
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device not returned to manufacturer.
 
Event Description
The manufacturer received information regarding a dreamstation device.The patient alleges can not sleep at night and dry mouth due to the device starts and stops during the night.The device is plugged directly into a wall outlet.The patient stated the device will not stay on all night despite being in fixed mode.The patient reported the heater plate "will get so hot you can't touch it." the patient detached humidifier from device and checked pins on the back of the device.They all appear normal, straight, and the pins are clean.The receiving port on the humidifier was clean and open.There was no report of flames, smoking, burning, melting or warping.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key19178133
MDR Text Key341061994
Report Number2518422-2024-22270
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUDSX500S11F
Device Catalogue NumberUDSX500S11F
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/31/2024
Initial Date FDA Received04/25/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/03/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-