• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. REAMER HANDLE OFFSET; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. REAMER HANDLE OFFSET; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 212760
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2024
Event Type  malfunction  
Event Description
Missing screw cross thread on impactor pitted handle.
 
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event Description
Missing screw cross thread on impactor pitted handle.
 
Manufacturer Narrative
An event regarding thread damage involving a mako reamer handle was reported.The event was not confirmed because the product was not available for inspection.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: a product history review is not required, as no manufacturing specific issue has been alleged.At stryker joint replacement all devices/product are manufactured through validated equipment and inspected by trained representatives through stryker¿s quality management system.Complaint history review: a complaint history review and trend detection is not required as this is a pm.Monitoring will be continued under the current quarterly trend review.Conclusions: no patient was involved and no actual or potential patient harm existed for the alleged event.The alleged failure mode was not confirmed because the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAMER HANDLE OFFSET
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer (Section G)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer Contact
sanjana talathi
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19179515
MDR Text Key341003682
Report Number3005985723-2024-00052
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00848486032142
UDI-Public00848486032142
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141989
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number212760
Device Lot Number5588827
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2024
Initial Date Manufacturer Received 04/01/2024
Initial Date FDA Received04/25/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-