• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer reporting a watchdog test failed message occurred on the v60 ventilator at power on.The device was not in clinical use.There was no report of harm.There was no patient involvement or impact.The device did not meet specification for intended use and was removed from service.A philips remote service engineer (rse) evaluated the issue with the biomedical engineer (bme) and verified diagnostic code 100b indicating the watchdog test failed in the device event log.A philips authorized service provider (asp) evaluated the device and confirmed the reported issue.The asp determined part replacement was necessary and replaced the gas delivery system (gds), motor control (mc) printed circuit board assembly (pcba) and the central processing unit (cpu) pcba.The device was operational and returned to service after repairs were completed.The investigation concludes that no further action is required at this time.
 
Manufacturer Narrative
H10: g1 phone number: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19182552
MDR Text Key341061438
Report Number2518422-2024-22418
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/09/2024
Initial Date FDA Received04/25/2024
Date Device Manufactured08/07/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-