U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60 V60PLUS VENTILATOR
Device Problem Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/15/2024
Event Type  malfunction  
Manufacturer Narrative
H10: g1 phone number (b)(6).
 
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator would not go into standby mode, and would not stay in standby mode.It was unknown how the issue was found.No patient or user harm reported.A philips remote service engineer (rse) noted that the customer's v60 ventilator would not go into standby mode and would not stay in standby mode.The customer reported that the device has software 3.00, and the high flow therapy (hft) option installed.It was noted that the air inlet filter was fairly dirty.The rse advised the customer that either the flow sensor assembly or gas delivery system (gds) assembly would need to be replaced to correct issue.The rse also discussed the management of the air inlet filter.The customer acknowledged and will order preferred part.The investigation is ongoing.
 
Manufacturer Narrative
H10: insufficient information is available to determine the resolution of the event.Multiple good faith efforts (gfe) were performed on 01may2024, 10may2024, and 15may2024 for further details regarding the event; however, no response was provided.No further details could be obtained regarding the event, and it could not be determined if the customer's issue was resolved or if the device had been repaired.No parts were returned for failure investigation.In the event that parts are returned or if new information is provided, the complaint will be reopened to update the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19183416
MDR Text Key341062189
Report Number2518422-2024-22471
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60 V60PLUS VENTILATOR
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/15/2024
Initial Date FDA Received04/25/2024
Supplement Dates Manufacturer Received05/23/2024
Supplement Dates FDA Received05/23/2024
Date Device Manufactured10/16/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-