RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Inadequate User Interface (2958)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/10/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had a touchscreen that was failing calibration.There was no patient involvement when the issue occurred.There was no patient or user harm reported.During troubleshooting, the remote service engineer (rse) provided the customer with the latest version of the service manual and the specific section for the repair of the reported issue.The rse also provided the customer with the part number for a replacement touchscreen.
 
Manufacturer Narrative
Insufficient information is available to determine the resolution of the event.Multiple good faith efforts (gfe) were performed for further details regarding the event; however, no response was provided.No further details could be obtained regarding the event, and it could not be determined if the customer's issue was resolved or if the device had been repaired.No parts were returned for failure investigation.In the event that parts are returned or if new information is provided, the complaint will be reopened to update the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key19184771
MDR Text Key341141191
Report Number2518422-2024-22527
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/10/2024
Initial Date FDA Received04/25/2024
Supplement Dates Manufacturer Received05/10/2024
Supplement Dates FDA Received05/14/2024
Date Device Manufactured03/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-