Correction: d1, d4: unique identifier (udi) #.H4: the lot was manufactured between april 2, 2023 and april 4, 2023.H11: the actual device and two photographs of the sample were received for evaluation.Visual inspection of photos observed a leak at the port 2 spike port bonding area.Visual inspection of returned sample did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing of actual sample was performed and a leak was identified from the spike port bonding area.The reported condition was verified.The cause of the condition could not be determined; however, the most probable cause was due to inadequate or lack of cyclohexanone being applied to the spike port cap tube when it was inserted into the spike port.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|