SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VH
Device Problem Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/03/2024
Event Type  malfunction  
Event Description
It was observed that during the device inspection, the videoscope distal end had foreign material near the objective lens.There were no reports of patient involvement.
 
Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The device evaluation found foreign material at the distal end near the objective lens, as well as a leak in the bending section cover.Based on the results of the investigation, it is likely that since the leakage failure occurred at the bending section cover, reprocessing could not be completed and foreign material remained at the distal end.However, a definitive root cause cannot be identified.A device history review revealed no issues that could have caused or contributed to the reported issue.This issue is addressed in the instructions for use (ifu): chapter 3 compatible reprocessing methods; chapter 4 reprocessing workflow for endoscopes and accessories; chapter 5 reprocessing the endoscope (and related reprocessing accessories); chapter 6 reprocessing the accessories; chapter 7 reprocessing endoscopes and accessories using an aer/wd.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19188786
MDR Text Key341356776
Report Number3002808148-2024-03899
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170411250
UDI-Public04953170411250
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K221683
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberCYF-VH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/18/2024
Initial Date FDA Received04/25/2024
Supplement Dates Manufacturer Received05/13/2024
Supplement Dates FDA Received05/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/15/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-