PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
|
Back to Search Results |
|
Model Number 15 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
A third party service agent evaluated the customer's device and was unable to duplicate the reported issue.Proper device operation was observed through functional and performance testing and the device was, subsequently, returned to the customer for use.
|
|
Event Description
|
The customer contacted stryker to report that their device had unexpectedly lost power.As a result, defibrillation would not be available, if needed.There was no patient use associated with the reported event.
|
|