• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS EARLYVUE VS30 VITAL SIGNS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS EARLYVUE VS30 VITAL SIGNS MONITOR Back to Search Results
Model Number 863359
Device Problem No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2024
Event Type  malfunction  
Event Description
It was reported there was a loudspeaker failure and no sound was being emitted from the device.It is unknown if the device was in clinical use at the time of the event, no adverse event or patient harm was reported.
 
Manufacturer Narrative
The device was sent for bench for repair.The following functional tests were performed: both speakers of the unit are exchanged.Issue was resolved by replacing the speakers.The client configuration is maintained and standard verification tests performed checked okay.Based on the information available and the testing conducted, the cause of the reported problem was a defective speaker.The reported problem was confirmed.The investigation concludes that no further action is required at this time.Reporting institution phone (b)(6) reporter phone (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EARLYVUE VS30 VITAL SIGNS MONITOR
Type of Device
EARLYVUE VS30 VITAL SIGNS MONITOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
deborah currlin
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key19189765
MDR Text Key341578374
Report Number1218950-2024-00300
Device Sequence Number1
Product Code DSJ
UDI-Device Identifier00884838091412
UDI-Public00884838091412
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K190624
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number863359
Device Catalogue Number863359
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/05/2024
Initial Date FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/13/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-