Model Number ARD568805906
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event Description
On 23rd april, 2024 getinge became aware of an issue with one of surgical lights - volista access 600.During maintenance it was found that dust cover (part catalog number: hm56053311) from spring arm was missing.There was no injury reported, however, we decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA ACCESS
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key19190023
MDR Text Key341120023
Report Number9710055-2024-00299
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodePO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD568805906
Device Catalogue NumberARD568805906
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/23/2024
Initial Date FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2020
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-